L-POR Electroporator

L-POR Electroporator
Pioneering Technology, First of Its Kind

Introducing the world’s first commercially available high-frequency electroporator—and the only device capable of generating asymmetrical bipolar pulses.

Revolutionary Precision, Tailored to You

Experience next-level innovation with our groundbreaking device—featuring 9 adjustable pulse parameters and 3 dynamic pulse modes.

Unlock New Frontiers in Electroporation Research

Its precision and versatility make it a powerful tool for researchers pushing the boundaries of science.

Laboratory High Frequency Electroporator

Streamlined for Everyday Lab Use
  • Easy To Use
  • Portable
  • Intuitive User Interface
Advanced Pulse Control, Redefined
  • 3 Pulse Modes
  • High Frequency
  • High Rise/Fall Time
Built for Control, Protected for Performance
  • External Trigger
  • Short Circuit Protection
  • Output Current 70 A

Watch Our Latest Video of L-POR Electroporator

L-POR high-frequency electroporator pulse specifications

L-POR electroporation device is a revolutionary device with 9 adjustable pulse parameters (positive/negative voltage range, positive/negative pulse duration, pause between positive and negative pulse, number of pulses in a burst, number of bursts and pulse/burst repetition rate) and 3 pulse modes (monopolar, bipolar symmetric and bipolar asymmetric).

It is perfectly suited for laboratory research as it enables completely new experiments in the electroporation field. It is the first commercially available high-frequency electroporator and the first device that also generates asymmetric bipolar pulses.The user interface is graphical in design to allow input of complex electroporation pulse parameters and control of measured signals.

The device features an integrated external trigger input and built‑in current protection, allowing continuous pulse delivery at currents up to 70 A without interruption.

L-POR Pulse Specifications Graph

 

PULSE PARAMETER MIN MAX
positive voltage range (Ap) 50 V 1400 V
negative voltage range (An) 50 V 1400 V
positive amplitude duration (Tp) 500 ns 10 ms
negative amplitude duration (Tn) 500 ns 10 ms
pause between positive and negative pulse (PAUSEpn) 500 ns 10 s
number of pulses in a burst (NrP) 1 1000
number of bursts (NrB) 1 1000
pulse repetition rate (PRR) 100 mHz 1 MHz
burst repetition rate (BRR) 100 mHz 100 Hz
pause between pulses (PBP) 500 ns 10 s
pause between bursts (PBB) 10 ms 10 s

Voltages can be set in 1 V increments. Durations can be set in 10 ns increments.