In Vivo PFA Ready
Developed for advanced laboratory and in vivo PFA research, L-POR 3 kV provides the high-voltage pulse generation and flexibility required for experimental pulsed field ablation protocols.
High-Voltage Pulse Generation up to 3 kV
Generate precisely controlled monopolar or bipolar pulses up to 3000 V, with pulse durations starting from 500 ns and a wide range of adjustable pulse parameters.
ECG-Synchronized Pulse Delivery
ECG-trigger compatible with R-wave synchronization, enabling precise synchronization of high-voltage pulse delivery with cardiac activity.
Laboratory High-Voltage High-Frequency Electroporator
L-POR 3 kV High-Frequency Electroporator Pulse Specifications

L-POR 3 kV is an advanced high-frequency electroporation device with 7 adjustable pulse parameters and 2 pulse modes: monopolar and bipolar. With an output voltage range from 100 V to 3000 V, pulse durations starting from 500 ns, and pulse repetition rates of up to 1 MHz, the device provides a highly flexible platform for advanced electroporation and pulsed field ablation research.
The device is designed for laboratory In Vivo PFA research, enabling the development and optimization of a wide range of experimental pulse protocols. Its graphical user interface allows intuitive control of complex pulse parameters and storage of user-defined protocols. An integrated HDMI output enables connection to an additional external display for convenient monitoring and visualization.


For easier experimental measurements, L-POR 3 kV is supplied with a dedicated measurement box that simplifies voltage and current monitoring during pulse delivery. The measurement box includes an integrated 1:10 voltage divider, allowing convenient connection to measurement equipment and easier observation of the delivered pulse waveform.
L-POR 3 kV features an integrated external trigger input and supports ECG-triggered pulse delivery with R-wave synchronization, allowing precise synchronization of high-voltage pulses with cardiac activity. Built-in current protection supports pulse delivery at currents of up to 70 A, providing additional protection during demanding experimental protocols.

| PARAMETER | MIN | MAX |
|---|---|---|
| Voltage range (A) | 100 V | 3000 V |
| Amplitude duration (T) | 500 ns | 10 ms |
| Pause between positive and negative pulse (PAUSEPN) | 500 ns | 10 ms |
| Number of pulses in a burst (NrP) | 1 | 1000 |
| Number of bursts (NrB) | 1 | 1000 |
| Pulse repetition rate (PRR) | 100 mHz | 1 MHz |
| Burst repetition rate (BRR) | 100 mHz | 100 Hz |
| Pause between pulses (PBP) | 500 ns | 10 s |
| Pause between bursts (PBB) | 10 ms | 10 s |
| Total treatment time duration | 500 ns | 40 ms |
| Total treatment burst duration | 500 ns | 80 ms |
Voltages can be set in 1 V increments. Durations can be set in 10 ns increments.